Application scope
Pharmaceutical industry, food industry, cosmetic industry, microculture, biotechnology &environmental simulation, etc
AIKANE constant temperature &humidity chamber can provide constant climatic conditions for stability test and the long-term stable climatic conditions shall comply with various standards and all requirements of ICH (International Council for Harmonization) guideline Q1A (R2) for new crude drug and preparation stability test.
了解更多
-The 7’’true color touch panel, smart operating system and high-performance processor contribute to easy and simple operation.
-Fuzzy PID control and programmed multi-section parameter setting.
-Precise timing function with week, day and hour modes.
-With the quick upgrade system, system program can be directly upgraded with USB.
-The system has automatic storage function and can store data of 3 years.
-Real-time check of temperature and humidity data or curve (for the last 8 to 10 hours).
-With event record function, 16,000 data can be recorded (including ON, OFF, menu parameter setting, fault alarm, etc.) and directly exported with USB.
-Imported Danfoss (SECOP) compressor and German EBM condenser fan.
-German HERAEUS temperature sensor and Austrian E+E capacitive transducer as the humidity sensor.
-Data monitoring standard access port (Ø50 mm) on the right, with silica gel plug.
-With CFC/HFC-free polyurethane foam as insulation material of the chamber, excellent thermal insulation and anti-condensation performance is realized.
-Samples can be observed through the inner glass door, thus avoiding affecting temperature and humidity due to door opening.-Flexible stainless steel -shelf system is easy to take out for cleaning and shelves have no sharp edges.
-Stainless steel liner (1.4301 / ASTM 304) and round corner design facilitate cleaning.
-Screensaver and coded lock prevent accidental operation.
-USB and RS485 interfaces.
-Chinese and English menu switch.
-When the temperature and humidity are set in the frostless range, the compressor works and the evaporator has no frost.
-When humidity is not needed, it can be set to zero and the equipment operates in the temperature mode; the control system can still ensure humidity inside the chamber.
-Energy saving: the energy saving effect is between 10% and 60% as the temperature and humidity settings change; water saving: the results change according to the temperature and humidity settings; for instance, a bottle of water can ensure operation of 30 to 60 days at the temperature of 25 °C and humidity of 60%. They jointly shorten working time of heating and humidification pipes, thereby extending their service life.
Technical parameter
Product model | DNK 250 | DNK 460 | DNK 760 |
Control mode/program | PID fuzzy logic control, common/program fuzzy operation | ||
Temperature control range | -10 °C~80 °C (-10 °C~90 °C/-40 °C~90 °C can be customized) | ||
Humidity control range | 15~95%RH/ ±1.5%RH | ||
Temperature fluctuation | ±0.1 °C | ||
Temperature uniformity at 25 °C | ±0.5 | ±0.5 °C | ±0.8℃ |
Humidity fluctuation | ±1.5RH | ||
Overall size (mm) W*H*D | 750*1,578*837 | 975*1,848*887 | 1,480*1,848*987 |
Liner size (mm) W*H*D | 580*980*440 | 795*1,180*490 | 1,300*1,180*500 |
Net weight | Approx. 196KG | Approx. 270KG | Approx.340KG |
Effective volume | 250L | 460L | 764L |
Total power of heating and humidification | 2200W | 2500W | 3600W |
Refrigerating power • refrigerant | 250W.R134A | 550W.R404A | 700W.R404A |
Temperature and humidity sensors | German HERAEUS temperature sensor and imported E+E humidity sensor | ||
Refrigerating method | Smart adjustment of compressor refrigerating capacity between 10% and 100%; large-scale frostless operation of refrigeration evaporation &significant reduction of energy and water consumption. | ||
Working temperature | 5 °C-35 °C; recommended working temperature is 10 °C-30 °C, in order to obtain optimal performance. | ||
Supply voltage | AC-220 V 50/60 HZ | ||
Safety device | Auto start, auto stop, power ON recovery, timing run, compressor over-voltage protection, over-temperature alarm, electric leakage protection, water shortage protection, door open protection, sensor anomaly alarm | ||
(Standard) tray | 3 layers (addable) | 4 layers (addable) | 4 layers (addable) |
* Performance parameter testing under idle conditions: 25 °C ambient temperature
* Product appearance parameters may be changed without notice and product appearance may show deviations due to photography, printing and other reasons.
Tip: Please carefully read the instruction manual before use.